M&C TechGroup SP2000-H320_S Data sheet User Manual

Gas sample probe series sp, Special features, Application

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

2.6

Embracing Challenge

ƒ

ƒ

Special probe behind DENOX (SCR)

ƒ

ƒ

Heated to 320 °C

ƒ

ƒ

No salt formation in the heated filter
part

ƒ

ƒ

Condensate vessel in the gas outlet with
glass globe filling for extension of the
reaction surface

ƒ

ƒ

Optional heated condensate vessel

ƒ

ƒ

ntegrated peristaltic pump

ƒ

ƒ

Connection for test gas feeding

ƒ

ƒ

Easy maintenance and operation

Special Features

Gas Sample Probe Series SP®

Version SP2000-H320/S heated to 320°C
with separator vessel

SP2000-H320/S

Application

M&C has developed a special sampling tech-

nique for continuous gas sampling of waste

gas in DENOX plants (SCR) where NH

3

is

added to the flue gas in order to reduce the

NOx content. This new sampling technique

has also proved to be suitable for processes

with very high pollutant concentrations.
In these applications, a big problem

is the measurement of NOx, SO

2

and

O

2

-concentrations. With temperatures of <

300 °C, ammonium salts are produced due to

the chemical reaction of NH

3

and SO

2

/SO

3

in

the flue gas.
This salification blocks up filters and sample

lines in a short time.
The special M&C gas sample probe SP2000-

H320/S represents a good solution for these

problems. In order to prevent a blocking

due to salification, the probe temperature is

adjusted above 300 °C.

Description

The M&C gas sample probe SP2000-H320/S is

based on the standard sample probe.
The gas sample probe SP2000-H320/S is tem-

perature controlled via an integrated capillary

sensor thermostat adjustable from 50 to 320

°C and including over temperature limiter and

low temperature alarm.
As an option, the gas sample probe is availa-

ble with a FeCu-Ni thermoelement instead of

the thermostat controller. For this version, an

external temperature controller is necessary.
Due to the modular design the M&C gas sam-

ple probe SP2000-H320/S can be equipped

with the wide range of different sample tubes

and pre-filters respectively options.
At the sample gas outlet of the probe, the

gas passes a heated adapter to a non-heat-

ed condensate vessel of glass. It is filled

with glass balls to extend the surface for

the salification. The salt deposits and can

be washed out with the condensate.

A peristaltic pump SR25.1G removes the con-

densate with the solved ammonium salts.
The temperature of the vessel is higher than

the ambient temperature because of the hot

gas stream and the heated adapter. Therefore

a loss of measured components is negligible

because of warm condensate. In case of

DENOX application with a small content of

NH

3

(normally only a few ppm) it is possible to

analyse SO

2

and NOx without great losses (only

some ppm which normally can be neglected).

To determine the loss, it is possible to give

test gas via the probe to the analyser(s).

A measuring fault can be found and cali-

brated.
As an option the vessel can be heated as

well to suppress chemical reactions of the

measured component below a defined tem-

perature.
To the sample outlet of the vessel a heat-

ed sample line 3/4-M for max. 200 °C

operating temperature can be connected.