M&C TechGroup SP34-H for portable gas sample probe PSP4000-H.. Data sheet User Manual
Electrically heated sample probe tube series sp, Special features, Application
M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
2.19
Embracing Challenge
Completely heated sample tube
Unheated pre-filters or extensions pos-
sible
Easy mounting
With thermostat, thermocouple Fe-CuNi
or PT100
Special Features
Electrically heated sample probe tube
series SP®
version SP34-H for portable gas sample probe PSP4000-H..
SP34-H
Application
The electrically heated M&C sample probe
tube SP34-H is used in extracting portable
sampling systems to avoid cooling and con-
densation of sample in the insitu tube from
the sample point to the heated sample probe
PSP4000-H.
Description
The standard length of electrically heated
M&C sample tube SP34 is 1000 mm. (Other
length on request)
The sample tube has G3/8” threads on both
ends. With the outside thread it will be
screwed into the inlet of the sample probe
PSP4000-H. Into the inside thread at the other
end, a non heated sample tube or prefilter
can be fixed.
The electrical heater is inside of a double
tube system, completely separated from the
process.
The temperature is controlled by an integrat-
ed capillary thermostat, built in the terminal
box. The temperature is adjustable in the
range of 100-180 °C. An external temperature
controller is not necessary. For the version
with thermocouple or PT100 an external tem-
perature controller is necessary.
Technical specifications and illustrations are without
obligation, subject to modifications. 05.97/06.06