Technical data – M&C TechGroup VC-2-SL Data sheet User Manual
Page 3

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
6.8
Technical specifications and illustrations are without
obligation, subject to modifications. 09.96/06.06
Pre-cooling unit version VC-...
VC-1
VC-1-SL
VC-2-L
VC-2-SL
Part No. 230V 50-60Hz
03K1000
03K3000
03K4000
03K5000
Part No. 115V 50-60Hz
03K1000
03K3000a
03K4000a
03K5000a
Jet-Stream heat exchanger out of Duran glass EC-G
1x
1x
2x
2x
Forced ventilation
no
yes
yes
yes
Autom. condensate removal with SR25.1
1x*
2x*
Sample gas connection
2x GL18-6
2x GL18-6
4x GL18-6
4x GL18-6
Condensate connection
1x GL25-12
1x DN4/6
2x GL25-12
2x DN4/6
Power consumption
25 VA
20 VA
30 VA
Gas flow rate recommended, (other flow rates possible)
1x 250 Nl/hr
1x 250 Nl/hr
2x 250 Nl/hr
2x 250 Nl/hr
Weight approximately
3,5 kg
7,5 kg
9 kg
11 kg
Sample gas pressure
max. 3 bar g
∆P per heat exchanger
4 mbar at 300 Nl/hr
Stagnant space per heat exchanger
approx. 70 ml
Sample inlet temperature
max. 180 °C
Ambient temperature
+2 °C to +45 °C
Storage temperature
0 °C to +55 °C
Electrical connection
Terminals max. 2,5 mm
2
, PG11 cable gland
Electrical standard / Case protection
EN 61010 / IP22-EN60529
Method of mounting / Ready for working
wall mounting / immediately
Material of sample contacting parts
Duran glass, PTFE
Duran glass, PTFE,
PVDF, Novoprene
Duran glass, PTFE
Duran glass, PTFE,
PVDF, Novoprene
Technical Data