5.1

Technical specifications and illustrations are without

obligation, subject to modifications. 08.99/06.06

*Standard, others to be indicated along with order, intermediate values possible. 1) approx. at 3 bar dilution gas behind pressure controller. 2) -5% not possible for orifice „g“. 3) with injec-

tor version I. Further technical data see leaflet SP2000, 2-1.1a.

Please set an (a) after the part no. for dilution probe with power 115V 60Hz, mounting flange ANSI 3" 150lbs or for connectors with inches dimensions.

Part no. 20 S 4300 : compl. set critical orifice a-g , injector orifice II , orifice seals

control panel -S, -S1 incl. wall mounting housing optional available: -S-G, -S1-G.

Other versions on request.

Technical Data

Part No.

Type

M&C gas sample dilution probe SP2000H/DIL .... with orifice ”e“ for dillution ratio 30 : 1 standard

20 S 4002

SP2000H/DIL

dilution probe 180 °C without ball valve upstream of the filter element

20 S 4102

SP2000H320/DIL

dilution probe 320 °C without ball valve upstream of the filter element

20 S 4005

SP2000H/DIL-VA

dilution probe 180 °C with heated manual operated ball valve upstream of the filter element

20 S 4022

SP2000H/DIL-B

dilution probe SP2000H/DIL with bypass injector and bypass gas to the vent

20 S 4024

SP2000H/DIL-BR

dilution probe SP2000H/DIL with bypass injector and bypass recirculation

20 S 4026

SP2000H/DIL-VA-B

dilution probe SP2000H/DIL-VA with bypass injector and bypass gas to vent

20 S 4030

SP2000H/DIL-VA-BV

dilution probe SP2000H/DIL-VA with bypass-needle valve and bypass gas to vent

20 S 4200

SP2000H/DIL-A

option: 1 pressure controller, 2 gauges, mounting set incl. connector set for dilution probe

20 S 4210

SP2000H/DIL-A1

option: 2 pressure controllers, 3 gauges, mounting set incl. connector set for dilution probe with bypass

20 S 4250

SP2000H/DIL-S

option: control panel with 1 pressure controller, 2 gauges, flowmeter, 2 ball valves

20 S 4260

SP2000H/DIL-S1

option: control panel with 2 pressure controllers, 3 gauges, flowmeter, 2 ball valves

SP2000H/DIL-B* -BR*

SP2000H/DIL-VA-B

SP2000H/DIL

Dilution gas

measuring point

sample gas out

Underpressure

180° C

Test gas

SP2000H320/DIL

Underpressure
measuring point

Test gas

320°C

Dilution gas

sample gas out

not present

with type *

-B

-BR

Bypass gas in

*

Sample gas Dilution gas

Underpressure
measuring point

180°C

Sample gas out

SP2000H/DIL-VA-BV

Underpressure
measuring point

Sample gas

180°C

Dilution gas

Sample gas out

Bypassgas out

Function Diagrams

Series SP

®

Version gas sample dilution probe SP2000H/DIL ...

Insitu sample tubes and pre filters optional on request

see leaflet 2-1.1a; 2-1.5.1; 2-1.6.1; 2-1.9.5

Dilution rates with the critical orifices ”a“ - ”g“

3)

a = 500

b = 200

c = 100

d = 50

e = 30*

f = 20

g = 10

: 1

Sample flow rate depending on the critical orifices ”a“ - ”g“

a = 1,4

b = 2,7

c = 5,5

d = 11

e = 19*

f = 28

g = 55

l/hr

1)

Possibility to adapt the dillution factror

with dillution gas pressure-adjustment -5% to +30%

2)

Dillution gas flow rate with injector version I or II

I: 480 - 600Nl/hr, optional for higher dilution rates II: 1800-3000 Nl/hr

Dilution gas pressure on inlet of pressure controller

min. 4,5 bar g, max. 16 bar g

Bypass injector /B: gas pressure-gas flow rate-sample gas flow rate

at approx. 2 bar g - injector gas approx. 300 l/hr - sample gas approx. 150 l/hr

Process pressure

0,9 up to 2 bar abs.

Fault caused by process temperature variations

operation independent from process temperature

Fault caused by process under- or overpressure

no fault as long as the differential pressure ∆P at the dilution unit is >0,5 bar g and test gas
is given to the probe under process conditions

Fault caused by atmospheric pressure variations

<1% with a variation of 50 mbar

Materials in contact with the sample gas

stainless steel 316 / 316Ti, quartz glass, FPM, graphite

Weight

approx. 20 kg

SP2000H/DIL-VA