Technical data, Sp2000h/dil-va-bv, Function diagrams – M&C TechGroup SP2000H320_DIL... Data sheet User Manual
Page 3: Sp2000h/dil-va

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
5.1
Technical specifications and illustrations are without
obligation, subject to modifications. 08.99/06.06
*Standard, others to be indicated along with order, intermediate values possible. 1) approx. at 3 bar dilution gas behind pressure controller. 2) -5% not possible for orifice „g“. 3) with injec-
tor version I. Further technical data see leaflet SP2000, 2-1.1a.
Please set an (a) after the part no. for dilution probe with power 115V 60Hz, mounting flange ANSI 3" 150lbs or for connectors with inches dimensions.
Part no. 20 S 4300 : compl. set critical orifice a-g , injector orifice II , orifice seals
control panel -S, -S1 incl. wall mounting housing optional available: -S-G, -S1-G.
Other versions on request.
Technical Data
Part No.
Type
M&C gas sample dilution probe SP2000H/DIL .... with orifice ”e“ for dillution ratio 30 : 1 standard
20 S 4002
SP2000H/DIL
dilution probe 180 °C without ball valve upstream of the filter element
20 S 4102
SP2000H320/DIL
dilution probe 320 °C without ball valve upstream of the filter element
20 S 4005
SP2000H/DIL-VA
dilution probe 180 °C with heated manual operated ball valve upstream of the filter element
20 S 4022
SP2000H/DIL-B
dilution probe SP2000H/DIL with bypass injector and bypass gas to the vent
20 S 4024
SP2000H/DIL-BR
dilution probe SP2000H/DIL with bypass injector and bypass recirculation
20 S 4026
SP2000H/DIL-VA-B
dilution probe SP2000H/DIL-VA with bypass injector and bypass gas to vent
20 S 4030
SP2000H/DIL-VA-BV
dilution probe SP2000H/DIL-VA with bypass-needle valve and bypass gas to vent
20 S 4200
SP2000H/DIL-A
option: 1 pressure controller, 2 gauges, mounting set incl. connector set for dilution probe
20 S 4210
SP2000H/DIL-A1
option: 2 pressure controllers, 3 gauges, mounting set incl. connector set for dilution probe with bypass
20 S 4250
SP2000H/DIL-S
option: control panel with 1 pressure controller, 2 gauges, flowmeter, 2 ball valves
20 S 4260
SP2000H/DIL-S1
option: control panel with 2 pressure controllers, 3 gauges, flowmeter, 2 ball valves
SP2000H/DIL-B* -BR*
SP2000H/DIL-VA-B
SP2000H/DIL
Dilution gas
measuring point
sample gas out
Underpressure
180° C
Test gas
SP2000H320/DIL
Underpressure
measuring point
Test gas
320°C
Dilution gas
sample gas out
not present
with type *
-B
-BR
Bypass gas in
*
Sample gas Dilution gas
Underpressure
measuring point
180°C
Sample gas out
SP2000H/DIL-VA-BV
Underpressure
measuring point
Sample gas
180°C
Dilution gas
Sample gas out
Bypassgas out
Function Diagrams
Series SP
®
Version gas sample dilution probe SP2000H/DIL ...
Insitu sample tubes and pre filters optional on request
see leaflet 2-1.1a; 2-1.5.1; 2-1.6.1; 2-1.9.5
Dilution rates with the critical orifices ”a“ - ”g“
3)
a = 500
b = 200
c = 100
d = 50
e = 30*
f = 20
g = 10
: 1
Sample flow rate depending on the critical orifices ”a“ - ”g“
a = 1,4
b = 2,7
c = 5,5
d = 11
e = 19*
f = 28
g = 55
l/hr
1)
Possibility to adapt the dillution factror
with dillution gas pressure-adjustment -5% to +30%
2)
Dillution gas flow rate with injector version I or II
I: 480 - 600Nl/hr, optional for higher dilution rates II: 1800-3000 Nl/hr
Dilution gas pressure on inlet of pressure controller
min. 4,5 bar g, max. 16 bar g
Bypass injector /B: gas pressure-gas flow rate-sample gas flow rate
at approx. 2 bar g - injector gas approx. 300 l/hr - sample gas approx. 150 l/hr
Process pressure
0,9 up to 2 bar abs.
Fault caused by process temperature variations
operation independent from process temperature
Fault caused by process under- or overpressure
no fault as long as the differential pressure ∆P at the dilution unit is >0,5 bar g and test gas
is given to the probe under process conditions
Fault caused by atmospheric pressure variations
<1% with a variation of 50 mbar
Materials in contact with the sample gas
stainless steel 316 / 316Ti, quartz glass, FPM, graphite
Weight
approx. 20 kg
SP2000H/DIL-VA
180° C
Test gas Dilution gas
measuring point
Sample gas out
Underpressure