Technical data – M&C TechGroup CSS-M_W Data sheet User Manual
Page 3

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany
[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111
14.4
Technical specifications and illustrations are without
obligation, subject to modifications. 04.04/06.06
Technical Data
Gas Conditioning Unit series CSS
®
version CSS-M
version CSS-M/W
Part No. for 230V 50Hz version
01G2000
01G2500
Part No. for 115V 60Hz version
01G2000a
01G2500a
Sample outlet dew point
range of adjustment: +2 °C ..... +15 °C, factory setting: +5 °C
Dew point stability
at constant conditions < ± 0,1 °C**
Sample inlet temperature
**max. 80 °C
Sample inlet dew point
**max.+50 °C
Gas flow rate / flow alarm
**max. 70 Nl/hr / no
**max. 70 Nl/hr / optional part no. 01G9805
Ambient temperature
**+5 °C up to +45 °C
Storage temperature
-25 °C up to +65 °C
Pressure
0,7 bar up to 1,4 bar abs.
Total cooling capacity **
max. 40 kJ/hr
Number of gas inlets
1
Number of gas outlets
1
Condensate outlet
1
Medium connections
tube connection 4/6 mm
Material of sample contacting parts
glass, PVDF, Novoprene
®
,FPM
Ready for operation
approx. 10 min.
Power supply
230V 50Hz or 115V 60Hz
Power consumption
max. 240VA
Fuse protection
4A t, 5x 20 mm
Electrical connection
Cold appliance plug with 2 m of cable
mains cable 3x1mm
2
,
signal cable 3x0,5mm
2
,
3m length
Status alarm, 1change over contact, poten-
tial free
no
contact rating 250VAC, 2A, summary alarm for gas flow, li-
quid and cooler temperature
Case protection
IP20 (DIN 40050. IEC 529)
Sheet steel housing, execution
portable, colour grey, RAL 7035
wall mounting, colour grey RAL 7035
Equipment dimensions
177 mm x 220 mm x 300 mm (H x W x D)
177 mm x 263 mm x 372,5 mm (H x W x D)
Electrical equipment standard
EN61010
Weight
approx. 11 kg
FPM
= Viton
®
PVDF = Polyvinylidenfluoride
**
Maximum values in technical datas must be rated in consideration of total cooling capacity at 25 °C ambient temperature and an outlet dew point of 5 °C.