beautypg.com

Technical data, Pressure difference and t90-time, Part number and weight – M&C TechGroup SP35-H... for gas sample probe SP2000-H... Data sheet User Manual

Page 2: Probe tube with pre-filter sp30-h1.1-v dimensions

background image

M&C TechGroup Germany GmbH • Rehhecke 79 • 40885 Ratingen • Germany

[email protected] • www.mc-techgroup.com • Fon +49 2102 935-0 • Fax +49 2102 935-111

2.15

Technical specifications and illustrations are without

obligation, subject to modifications. 12.97/06.06

Series SP

®

Type

Version SP35- .... Version SP30- ....
H1.1 H2 H1.1 H2 H1.1-V

Temperature sensor

Fe-CuNi

PT100 2

Fe-CuNi

PT100 2

Fe-CuNi

Temperature controller

optional, external

Probe tube length L1

175 mm

max. 2 m

max. 1 m

Sample temperature max.

550 °C

Operating temperature max.

320 °C

200 °C

320 °C

200 °C

320 °C

Pre-filter length 520 mm, ø 60 mm,
filter porosity 2µm, integrated and heated

V20-2/30

Sample gas inlet connection

G3/4"i DIN ISO 228/1

pre-filter with G1 1/2"i
DIN ISO 228/1

Dust loading

max. 2 g/m

3

>2 g/m

3

Probe tube volume

380 ml/m

420 ml

Sample pressure max.

5 bar g

Ambient temperature

-20 °C to +80 °C

Storage temperature

-30 °C to +90 °C

Ready for operation

2 h

Power supply

230V

AC

, (115V optional)

230/115V

AC

- switchable,

Heating capacity

200W

0,6m: 600W, 1m: 800W, 1,5m: 1200W, 2m: 1200W

Electrical connections

terminals, max. 2,5 mm

2

, 2x PG13,5 cable gland

Electrical standard

EN 61010, EN60519-1

Degree of protection

IP54 EN 60529

Mounting flange

DN65 PN6, Form B with 4 welded screws on both sides M 12x 40 mm

Material of parts in contact with the sample

stainless steel SS316Ti, Version SP30-H1.1/HC: Hastelloy C

1.4539, SS316Ti/SS316

Technical Data

∆P and T

90

at flow:

100

200

500

1000

Nl/h

∆P Pressure loss SP30-H... length 1m

< 1

< 1

< 1

< 1

mbar

∆P pressure loss SP30-H...-V
length 1m with new pre-filter V20-2/30

< 1

< 1

1,5

4

mbar

T

90

- time

14

7

< 3

< 2

sec

Pressure difference and T90-time

Version SP35-...

Version SP30- ...

execution:

length „L1“

175 mm

0,6 m

1,0 m

1,5 m

2,0 m

Part number

20S9320

20S9322

02S9023
02S9026
02S9023p

02S9025
02S9092
02S9025p

02S9026
02S9093
02S9026p

02S9024
02S9094
02S9024p

H1.1
H1.1/HC
H2

Weight

4,8 kg

6,4 kg

8,0 kg

9,8 kg

11,6 kg

Part number

02S9037

02S9038

H1.1-V

Weight

9,0 kg

10,6 kg

+ 02S9036

Standard, switchable

115V

AC

Part number and weight

210

120

Ø 130

20

∅ 160

14 40

M12

L1

Terminals

Ø 42,2

G3/4"

PG13,5

Temperature-Sensor

Power

Temperature of pre filter in dependence of controller

temperature at 20°C ambient temperature.

500

ш 64

210

120

ш 130

ш 160

14 40

M12

L1

Terminals

PG13,5

Temperature

Sensor

Power

Controller temperature

180°C

250°C

300°C

Temperature of pre filter

135°C

182°C

220°C

Probe tube SP30-H1.1/H2, SP35-H1.1/H2

Dimensions in mm

Probe tube with pre-filter SP30-H1.1-V

Dimensions

The corresponding temperature controller has to be ordered and mounted separatly. See leaflet 2-5.1.