beautypg.com

Table 5-2 storm water runoff sampling (continued) – Teledyne Refrigerated Sampler 6712FR User Manual

Page 97

background image

6712FR Refrigerated Sampler

Section 5 Extended Programming

5-9

35

ENABLE:

RAIN LEVEL FLOW

NONE

Select RAIN.

Note: All valid enable options will appear here, so your screen may differ.

36

ENABLE: RAIN

AND

OR

DONE

Select AND.

37

ENABLE: RAIN AND

LEVEL FLOW NONE

Select LEVEL.

38

“RAIN” SET POINT:

0.15 INCHES PER

__HOURS, __MINUTES

Enter 0.15 for inches, then enter 30 minutes.

39

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

Select ABOVE SET POINT.

40

RESET RAIN HISTORY?

YES

NO

Select YES.

41

“LEVEL” CONDITION:

SET POINT

RANGE

RATE OF CHANGE

Select SET POINT.

42

“LEVEL” SET POINT

0.25 ft

(0.001-30.000)

Enter 0.25.

43

ENABLED WHEN:

ABOVE SET POINT

BELOW SET POINT

Select ABOVE SET POINT.

44

ONCE ENABLED,
STAY ENABLED?

YES NO

Select YES.

45

SAMPLE AT ENABLE?

YES NO

Select YES.

46

PAUSE RESUME

1. HH:MM DD HH:MM DD
2. HH:MM DD HH:MM DD

CLEAR DONE

Select DONE.

Screen will say “End of Part A” then “Beginning Part B.”

47

UNIFORM TIME PACED

FLOW PACED

EVENT PACED

NONUNIFORM TIME

Select FLOW PACED.

Table 5-2 Storm Water Runoff Sampling (Continued)