6712FR Refrigerated Sampler
Section 8 Maintenance

8-6

8.4.3 Internal Battery

Warning

The internal lithium battery preserves the sampler’s memory
when power is removed. For an explanation of how to replace the
lithium battery, go to Section 8.8

.

Table 8-2 Resetting the Pump Counter

1

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

Select OTHER FUNCTIONS.

2

MAINTENANCE

MANUAL FUNCTIONS

PROGRAMMING STYLE

HARDWARE

Select MAINTENANCE.

3

SET CLOCK

PUMP TUBE ALARM
INTERNAL BATTERY

DIAGNOSTICS

Select PUMP TUBE ALARM.

4

1024000 PUMP COUNTS,

WARNING AT 1000000

The sampler displays this screen briefly to give you the current pump-count
information. The first line shows you the pump counts tallied since the
counter was last reset. The second line shows the current alarm setting.

5

1024000 PUMP COUNTS,

RESET PUMP COUNTER?

YES NO

To reset the counter to zero, select YES. Always reset the counter after
replacing a pump tube. Select NO when merely checking the current count.

6

WARNING AT 1000000

PUMP COUNTS

(1 - 99)00000)

If necessary, change the pump-count alarm setting by typing the first two
digits of the new setting. The sampler accepts entries between 1 and 99. For
example, to increase the count to 1,500,000, enter 15.

7

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

Press

STOP

to return to the main menu.

Table 8-3 Checking and Replacing the Internal Battery

1

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

Select OTHER FUNCTIONS.

2

MAINTENANCE

MANUAL FUNCTIONS

PROGRAMMING STYLE

HARDWARE

Select MAINTENANCE.

3

SET CLOCK

PUMP TUBE ALARM

INTERNAL BATTERY

DIAGNOSTICS

Select INTERNAL BATTERY.

4

WARNING!

REPLACE INTERNAL

BATTERY BEFORE

17-OCT-97

The sampler displays this screen briefly to show you the expiration date.