Section 4 Standard Programming
4-22
Figures 4-5, 4-6, 4-7, 4-8, and 4-9 show examples of reports con-
figured to show program settings, sampling results, combined
results, and a summary report.
(The Source and Error Codes Listed below refer to why a sample was taken and what happened while it was being
taken.)
Source Codes
*C
Calibrate sample
Sample volume delivered during calibration.
*D
Disable
Sample event at disable time.
*E
Enable
Sample event at enable time.
Ev
Event
Event-paced sample event.
*F
Flow
Flow-paced sample event.
*G
Grab sample
Sample volume delivered as a grab sample.
M
Command driven sample A command driven sample.
PH
Phone command
Phone command sample.
R
Resume
Sample event at resume time.
*S
Start
Sample event at start time.
Sw
Switch time
Sample event at switch time.
TO
Time Override
Sample event at bottle-switch time.
*T
time
Time-paced sample event.
Error Codes
*DJ
Distributor Jammed
The distributor jammed.
*IP
sample in progress
Sample in progress when report was printing.
*L
pump latch open
Unable to take sample because the pump band was opened.
*ND no distributor arm
There is no distributor arm or the stop is damaged.
*NL
No Liquid Detected
The sampler was unable to detect liquid.
LF
Not enough flow
Not enough flow for the minimum sample volume entered for samples with volumes
proportional to flow.
*NM No More Liquid
After the sampler detected liquid and while the sample was being taken, the liquid
detector stopped detecting liquid.
M
Bottle Full
The composite bottle is full.
*O
Sampler Shut Off
The sampler was turned off with the On-Off key during the sample event.
Ov
Probable Overflow
The sampler was directed to take a sample that would not fit into the bottle. Occurs
for flow paced, time override programs only.
*P
Power Failed
The power supply failed during the sample event. Note: Power failures during five
consecutive sample events results in a LOW BATTERY shut-down.
*PJ
Pump Jammed
The pump jammed.
*Sk
Sample Skipped
Sample not taken.
*US User Stopped
You pressed the Stop key during the sample event.
Table 4-3 Reports: Program Events, Source Codes, and Error Codes (Continued)