6712FR Refrigerated Sampler

Section 5 Extended Programming

5-7

11

PROGRAM MODULE?

YES NO

Select YES

12

MODE OF OPERATION

FLOWMETER

LEVEL ONLY

Select FLOWMETER

13

WEIR FLUME

DATA POINTS

EQUATION MANNING

FLOW-INSERT

Select FLOW-INSERT.

14

WEIR/ORIFICE TYPE

V-NOTCH

ROUND

Select ROUND.

15

FLOW INSERT SIZE:

6" 8" 10" 12"

Select 12.

16

NEW MODULE SETUP--

DOWNLOAD DATA NOW

OR LOSE ALL DATA!

DONE

Select DONE.
Note: Partition memory status message will be displayed briefly.

17

CURRENT LEVEL IS

___.___ ft.

ADJUST LEVEL TO

___.___ ft.

Press ENTER.

18

DATA STORAGE

INTERVAL IN MINUTES

1 2 5

10 15 30

Select 5. (If the data storage interval is different from the current program,
you will be informed that the storage interval has changed, and that the
stored data will be lost. Download stored data (if desired) and select DONE.

19

NUMBER OF BOTTLES:

1

2

4

8

12

24

Select 24.

20

BOTTLE VOLUME IS

1000 ml (300-30000)

Enter 1000.

21

SUCTION LINE LENGTH

IS 5 ft
(3-99)

Enter 5. The 6712FR calculates new pump tables.

22

AUTO SUCTION HEAD

ENTER HEAD

Select AUTO SUCTION HEAD.

Table 5-2 Storm Water Runoff Sampling (Continued)