6712FR Refrigerated Sampler
Section 2 Installation/Preparation

2-2

Figure 2-1 Mounting the Controller

2.1.3 Installing the

Distributor Arm and
Discharge Tube

The distributor arm is adjustable so that you can position the
discharge tube over the bottles of each kit. Each bottle kit also
uses a discharge tube of a specific length. Table 2-1 lists the
correct discharge tube lengths for each bottle kit. If you are using
bulk tubing to cut your own tubes, cut the ends of the tube
square, not at an angle, and cut the length accurately.

The two-piece distributor arm has four positions, marked on the
underside with the letters A, B, C, and D. Adjust the arm to the
correct position for the bottle kit being used. Table 2-1 shows the
correct position for each kit. Thread the discharge tube through
the spring and arm so that the tube protrudes from the arm

1

/

16

to

1

/

8

inch. If too much tube protrudes from the arm it can catch

on bottles and jam the distributor arm.

Attach the arm to the distributor shaft extension. Secure the arm
to the extension with the black plastic nut. Slide the free end of
the discharge tube onto the bulkhead fitting. After installing the
distributor arm and discharge tube, run the distributor to each
bottle position to ensure proper installation.

Multiple-Bottle Sampling

Wing Nut

Stop Arm

Distributor arm
Assembly

1/4-20 Set Screw

Distributor arm
Nut

Distributor Shaft

Extension

Mounting Rod

Flat Washer

Composite Sampling

Mounting Rod

Flat Washer

Wing Nut

below neck of bottle.

End of discharge tube

Distributor Arm