6712FR Refrigerated Sampler
Section 4 Standard Programming
4-29
CAUTION
Most Isco accessories and interfacing equipment are not certi-
fied for us in hazardous locations as defined by the National
Electrical Code. Check applicable guidelines, codes, and regu-
lations of federal, state, city, and county agencies. Refer to
Appendix C in the back of this manual for more safety informa-
tion.
Table 4-4 Programming Example With 730 Module Installed
1
MODULE INSERTED---
DOWNLOAD DATA NOW
OR LOSE ALL DATA!
DONE
With the sampler off, insert the module in the bay on the controller.
Turn the sampler on by pressing the
On/Off
key. Press
↵
.
Download stored data, if you want to keep it. Pressing DONE erases it and
advances to the next screen.
2
6712 SAMPLER
STANDARD PROGRAMMING
For HELP at any
screen press ? key.
This screen will now appear for 8 seconds.
3
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
The option PROGRAM will be blinking. Press
↵
.
4
SITE DESCRIPTION:
“FACTORY051“
CHANGE?
YES NO
The option NO will be blinking. Press
↵
.
For the purposes of this example, it isn’t necessary to change the description.
5
SELECT UNITS FOR
FLOW RATE:
cfs gps gpm Mgd
lps m3s m3h m3d
Select the desired units for the flow rate. For this example, choose cfs.
Press
↵
.
6
SELECT UNITS FOR
FLOW VOLUME:
cf gal Mgal
m3 lit
Select the desired units for the flow volume. For this example, choose Mgal.
Press
↵
.
7
PROGRAM MODULE?
YES NO
Select YES to program the module.
8
MODE OF OPERATION:
FLOWMETER
LEVEL ONLY
If you choose Flowmeter, proceed to screen #9 below.
If you choose LEVEL ONLY, proceed to screen #12.
9
WEIR FLUME
DATA POINTS
EQUATION MANNING
FLOW-INSERT
When you select Flowmeter, you will have several options for performing the
flow conversion.
For this example, choose Weir. Press
↵
.