4.12.1 Calibration Tips

• For the best results, calibrate the sampler after it’s

installed on site.

• Use a graduated cylinder for volume measurement.

• The sampler clears the previous calibration setting

when you reinitialize the sampler.

• You cannot calibrate while the sampler is running a

program.

• When the sampler delivers the sample, it runs through a

complete sampling cycle, using the current settings for
volume line rinses and retries. If it is a two part
program, the sampler uses the volume setting for part
A.

To calibrate:

1. Disconnect the pump tube from the bulkhead fitting.

2. Place the end of the tube over a collection container.

3. Follow the steps in Calibrating Sample Volumes in the

margin.

Reconnect the pump tube to the bulkhead fitting.

Calibrating Sample Volumes

1

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

•Select OTHER FUNCTIONS

2

MAINTENANCE

MANUAL FUNCTIONS

MODULE

HARDWARE

•Select MANUAL FUNCTIONS

3

GRAB SAMPLE

CALIBRATE VOLUME

OPERATE PUMP

MOVE DISTRIBUTOR

•Select CALIBRATE VOLUME

4

SAMPLE VOLUME:

___ ml (100-9990)

•Enter the desired volume

5

CALIBRATE VOLUME

PRESS

↵ WHEN READY!

6

PURGING

SUCTION LINE

7

PUMPING 200 ml

8

PURGING

SUCTION LINE

9

VOLUME DELIVERED:

____ml

•Measure the volume delivered

and enter the amount here.

10

200 ml

ARE YOU SURE?

YES NO

This screen appears when the
amount you entered in step 9
was more than twice, or less
than half, the programmed
volume.

• Select YES when the volume

delivered matches the volume
displayed on this screen.

• Select NO to enter the volume

delivered.