1 calibration tips, 1 calibration tips -16 – Teledyne Refrigerated Sampler 6712FR User Manual
Page 72
6712FR Refrigerated Sampler
Section 4 Standard Programming
4-16
4.12.1 Calibration Tips
• For the best results, calibrate the sampler after it’s
installed on site.
• Use a graduated cylinder for volume measurement.
• The sampler clears the previous calibration setting
when you reinitialize the sampler.
• You cannot calibrate while the sampler is running a
program.
• When the sampler delivers the sample, it runs through a
complete sampling cycle, using the current settings for
volume line rinses and retries. If it is a two part
program, the sampler uses the volume setting for part
A.
To calibrate:
1. Disconnect the pump tube from the bulkhead fitting.
2. Place the end of the tube over a collection container.
3. Follow the steps in Calibrating Sample Volumes in the
margin.
Reconnect the pump tube to the bulkhead fitting.
Calibrating Sample Volumes
1
RUN
PROGRAM
VIEW REPORT
OTHER FUNCTIONS
•Select OTHER FUNCTIONS
2
MAINTENANCE
MANUAL FUNCTIONS
MODULE
HARDWARE
•Select MANUAL FUNCTIONS
3
GRAB SAMPLE
CALIBRATE VOLUME
OPERATE PUMP
MOVE DISTRIBUTOR
•Select CALIBRATE VOLUME
4
SAMPLE VOLUME:
___ ml (100-9990)
•Enter the desired volume
5
CALIBRATE VOLUME
PRESS
↵ WHEN READY!
6
PURGING
SUCTION LINE
7
PUMPING 200 ml
8
PURGING
SUCTION LINE
9
VOLUME DELIVERED:
____ml
•Measure the volume delivered
and enter the amount here.
10
200 ml
ARE YOU SURE?
YES NO
This screen appears when the
amount you entered in step 9
was more than twice, or less
than half, the programmed
volume.
• Select YES when the volume
delivered matches the volume
displayed on this screen.
• Select NO to enter the volume
delivered.