6712FR Refrigerated Sampler
Section 4 Standard Programming

4-2

Use the arrow key to move between the two choices, feet or
meters. When the desired choice is blinking, press

↵. The default

unit of length for the Spanish option is meters.

4.3 Programming

Examples

The following table is an example of time-paced sampling.
Further details about some of the menu items are provided in
later sections.

Table 4-1 Standard Program: Sample Every 15 Minutes, One Sample Per Bottle

1

6712 SAMPLER

STANDARD PROGRAMMING

For HELP at any

screen press ? key.

Turn the sampler on by pressing the

On/Off

key. Press

.

This screen disappears on its own after 8 seconds.

2

RUN

PROGRAM

VIEW REPORT

OTHER FUNCTIONS

The option PROGRAM will be blinking. Press

.

3

SITE DESCRIPTION:

“FACTORY051“

CHANGE?

YES NO

The option NO will be blinking. Press

.

For the purposes of this example, it isn’t necessary to change the description.

4

NUMBER OF BOTTLES:

1 2 4 8 12 24

Select the number of bottles in your bottle kit by pressing either arrow key
until the correct number blinks. Press

. For this example, select 24.

5

BOTTLE VOLUME IS

1000 ml (300-30000)

Type the volume for the bottles in your kit. For this example, 1000 is correct,
so simply press

.

6

SUCTION LINE LENGTH

IS 10 ft

(3-99)

Type the length of the suction line, then press

.

If you change the length, the sampler will display a message,
“PLEASE WAIT!. . . GENERATING PUMP TABLES.”

7

TIME PACED

FLOW PACED

Because this example requires samples every 15 minutes, select TIME
PACED by pressing an arrow until the option TIME PACED blinks. Then,
press

.

8

TIME BETWEEN

SAMPLE EVENTS

0 HOURS, 15 MINUTES

Type 0 for hours and press

. Type 15 for minutes and press

.

Tip: Move back and forth between hours and minutes by pressing an arrow
key.

9

SEQUENTIAL

BOTTLES/SAMPLE
SAMPLES/BOTTLE

Because this program requires one sample in each bottle, select
SEQUENTIAL by pressing an arrow until the option SEQUENTIAL blinks.
Then, press

.

10

RUN CONTINUOUSLY?

YES NO

For this example, select NO by pressing an arrow until the option blinks.
Then, press

. Selecting YES allows the program to run indefinitely by

repeating the sample distribution. Continuous sampling assumes that filled
bottles are replaced with empty bottles at regular service intervals.