Fail output, Chart end output – Omega RD100B User Manual
Page 41
1-27
Functional Description
1
1.7
FAIL/Chart End Detection and Output Function
(/F1 Option)
FAIL Output
When a failure occurs in the CPU of the recorder, a relay contact signal (1 signal) is
output. The relay is energized when the CPU is normal and de-energizes when a CPU
failure occurs. Therefore, relay output is carried out also when the power is turned OFF
(including a power failure). This relay behavior cannot be changed.
• FAIL Relay Output
NO
C
NC
Normal
NO
C
NC
Failure
NO
C
NC
Power-OFF
NO: Normally Opened, C: Common, NC: Normally Closed
Chart End Output
When the remaining amount of chart paper falls to approximately 2 cm, recording is
stopped, and the relay contact signal (1 relay) is output. The relay is energized in this
case. This relay behavior cannot be changed.
NO
C
NC
Normal
NO
C
NC
Chart end
NO
C
NC
Power-OFF
NO: Normally Opened, C: Common, NC: Normally Closed