Omega RD100B User Manual

Page 181

background image

9-24

9.14 Changing the Channel Assignments of

Recording Pens (Pen Model)

Measurement channels and computation channels can be arbitrarily assigned to

recording pens.

Basic Setting mode cannot be entered when recording is in progress or when

computation is in progress on models with the computation function (/M1 option).

Procedure

Changing the Settings

1.

Hold down the

MENU

key for 3 seconds to enter Setting mode.

2.

Hold down the

and

keys simultaneously for 3 seconds to enter Basic

Setting mode.

3.

Press the

key or SHIFT +

key to select Math and then press the

key.

4.

Press the

key or SHIFT +

key to select Output pen and then press the

key.

5.

Set each item and press the

key.

Use the

key or SHIFT +

key to select values.

For the procedure on how to enter values or characters, see section 4.2.

If you press the ESC key in the middle of the operation, the settings entered up

to then are cancelled, and the display returns to a higher level menu.

6.

When the Setting complete screen appears, do either of the following:

To correct the setting, press the

key.

If you are done, press the ESC key.

Select the pen number and channel
number.

The new setting takes effect.

ESC/?

Pen 1
Setting complete

Basic=Math

Math=Output pen

Pen=1 CH=01

Applying the Changes and Returning to Operation Mode

Press the

key or SHIFT +

key to select End and then press the

key. Press

the

key to select Store and then press the

key. The changes are applied, and

the screen returns to Operation mode. If you select Abort and press the

key, the

changes are discarded, and the screen returns to Operation mode. Press the ESC key

to return to the Basic= screen.

Explanation

Output pen

Measurement channels and computation channels can be arbitrarily assigned to

recording pens.

The default values are as follows:

Pen No. 1 (red): Channel 01

Pen No. 2 (green): Channel 02

Pen No. 3 (Blue): Channel 03

Pen No. 4 (violet): channel 04