9-24
9.14 Changing the Channel Assignments of
Recording Pens (Pen Model)
Measurement channels and computation channels can be arbitrarily assigned to
recording pens.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key or SHIFT +
key to select Output pen and then press the
key.
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
To correct the setting, press the
key.
If you are done, press the ESC key.
Select the pen number and channel
number.
The new setting takes effect.
ESC/?
Pen 1
Setting complete
Basic=Math
Math=Output pen
Pen=1 CH=01
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.
Explanation
Output pen
Measurement channels and computation channels can be arbitrarily assigned to
recording pens.
The default values are as follows:
Pen No. 1 (red): Channel 01
Pen No. 2 (green): Channel 02
Pen No. 3 (Blue): Channel 03
Pen No. 4 (violet): channel 04