9-28
9.17 Setting the Procedure Taken When the
Computed Result Is in Error
This section explains the details of setting the procedure taken when the computed result
is in error.
Basic Setting mode cannot be entered when recording is in progress or when
computation is in progress on models with the computation function (/M1 option).
Procedure
Changing the Settings
1.
Hold down the
MENU
key for 3 seconds to enter Setting mode.
2.
Hold down the
and
keys simultaneously for 3 seconds to enter Basic
Setting mode.
3.
Press the
key or SHIFT +
key to select Math and then press the
key.
4.
Press the
key or SHIFT +
key to select Error data and then press the
key.
5.
Set each item and press the
key.
Use the
key or SHIFT +
key to select values.
For the procedure on how to enter values or characters, see section 4.2.
If you press the ESC key in the middle of the operation, the settings entered up
to then are cancelled, and the display returns to a higher level menu.
6.
When the Setting complete screen appears, do either of the following:
Press the
key to set other channels.
If you are done, press the ESC key.
Select the computation error procedure.
The new setting takes effect.
ESC/?
Select the input overflow procedure.
Math exception
Setting complete
Basic=Math
Math=Error data
Error=+Over
Over=Skip
Applying the Changes and Returning to Operation Mode
Press the
key or SHIFT +
key to select End and then press the
key. Press
the
key to select Store and then press the
key. The changes are applied, and
the screen returns to Operation mode. If you select Abort and press the
key, the
changes are discarded, and the screen returns to Operation mode. Press the ESC key
to return to the Basic= screen.