beautypg.com

Omega RD100B User Manual

Page 197

background image

11-5

Maintenance

11

Temperature Measurement When Using an RTD

Decade resistance box

The resistance of three lead wires must be equal.

+/A

/b

Input terminals

/B

Temperature Measurement When Using a Thermocouple

Thermocouple wires or TC extension wires

Copper wires

DC voltage standard

+

Input terminals

+

Thermocouple
wires

Copper
wires

(0°C standard temperature device)

RJC of TC input

As the measurement terminal of the recorder is generally at room temperature, the

actual output of the thermocouple is different from the values given on the

thermoelectromotive force table based on 0°C. The recorder performs compensation by
measuring the temperature at the input terminal and adding the corresponding

thermoelectromotive force to the actual output of the thermocouple. Therefore, when the

measurement terminal is shorted (equivalent to the case when the detector tip is 0°C),
the measured value indicates the temperature of the input terminal.

When calibrating the recorder, this compensation voltage (thermoelectromotive force of

0°C reference corresponding to the input terminal temperature) must be subtracted from
the output of the DC voltage standard before application. As shown in the figure, by

using the 0°C standard temperature device to compensate the reference junction at 0 °C,
you can input the thermoelectromotive force of 0 °C reference from the DC voltage
standard and perform the calibration.

11.4 Calibrating the Recorder