beautypg.com

Omega RD100B User Manual

Page 22

background image

1-8

• Delay High Limit Alarm (T)

An alarm occurs when the measured value remains above the alarm value for a

specified time period (alarm delay period).

• Delay Low Limit Alarm (t)

An alarm occurs when the measured value remains below the alarm value for a

specified time period (alarm delay period).

Delay High Limit Alarm Example (T is the specified delay)

Alarm release

Alarm occurrence

T

X1

X2

X3

X4

T1

Measured value

Alarm value

• Alarm does not occur at T1, because the time period is shorter than the specified

alarm delay period (T).

• The measured value exceeds the alarm value at time X2, and the alarm occurs at

time X3 at which the specified alarm delay period elapses (the time when the alarm

occurs is the time at X3).

• The measured value falls below the alarm value at time X4, and the alarm is

released.

Note

The alarm detection operation is reset when a power failure occurs. The operation

restarts after the power recovers.

If the alarm value is changed while a delay alarm is occurring, the alarm is released if the

new alarm value does not meet the alarm condition.

Alarm Hysteresis

Hysteresis can be specified to the values for activating and releasing the alarm. The

hysteresis applies only to high limit alarm (H) and low limit alarm (L). The hysteresis

width can be set in the range of 0.0% (Off) to 1.0% of the recording span in 0.1 steps.

The setting applies to all high limit alarms and low limit alarms. By default, the hysteresis

width is set to 0.5%.

Alarm
value

Alarm release

Measured value

Alarm occurrence

High limit alarm

Low limit alarm

Measured
value

Alarm release

Hysteresis
(1% or less)

Alarm value

Alarm occurrence

Setting alarms: Section 5.2

Setting the alarm delay function: Section 7.15 and 6.10

Setting the alarm hysteresis: Section 7.1

1.3 Alarms