beautypg.com

National Instruments Data Acquisition Device NI USB-621x User Manual

Page 83

background image

Chapter 6

Digital I/O

© National Instruments Corporation

6-3

NI USB-621x User Manual

Increasing Current Drive on USB-6210/6211/6215/6218 Devices

The total internal current limit for digital outputs and power drawn from the
+5 V terminals is 50 mA. You can increase this internal current limit by
supplying an external +5 V supply. Refer to the

+5 V Power as an Input

section of Chapter 3,

Connector and LED Information

.

Connecting Digital I/O Signals on USB-6210/6211/6215/6218 Devices

The DI and DO signals, P0.<0..7> and P1.<0..7> are referenced to D GND.
Digital input applications include receiving TTL signals and sensing
external device states, such as the state of the switch shown in the figure.
Digital output applications include sending TTL signals and driving
external devices, such as the LED shown in Figure 6-2.

Figure 6-2. USB-6210/6211/6215/6218 Digital I/O Connections

Caution

Exceeding the maximum input voltage ratings, which are listed in the

NI USB-621x Specifications, can damage the DAQ device and the computer. NI is not
liable for any damage resulting from such signal connections.

+5 V

LED

TTL Signal

+5 V

Switch

I/O Connector

D GND

USB-6210/6211/6215/6218 Device

P0.<0..3>

P1.<0..3>

Digital

Isolators

Isolation

Barrier

(USB-6215

and USB-6218

devices only)

When using a

USB-6215/6218,

you must connect

D GND and/or AI GND

to the local ground

on your system.