Routing signals to counter n hw arm input, Counter n internal output and counter n tc signals, Frequency output signal – National Instruments Data Acquisition Device NI USB-621x User Manual
Page 122: Routing frequency output to a terminal, Routing signals to counter n hw arm input -29, Frequency output signal -29, Routing frequency output to a terminal -29, Counter n internal output signal, Counter n tc signal

Chapter 8
Counters
© National Instruments Corporation
8-29
Routing Signals to Counter n HW Arm Input
Any of the following signals can be routed to the Counter n HW Arm input:
•
(USB-6210/6211/6215 Devices)
PFI <0..3>
•
(USB-6212/6216 Devices)
PFI <0..15>
•
(USB-6218 Devices)
PFI <0..3>, PFI <8..11>
•
AI Reference Trigger (ai/ReferenceTrigger)
•
AI Start Trigger (ai/StartTrigger)
Counter 1 Internal Output can be routed to Counter 0 HW Arm. Counter 0
Internal Output can be routed to Counter 1 HW Arm.
Some of these options may not be available in some driver software.
Counter n Internal Output and Counter n TC Signals
The Counter n Internal Output signal changes in response to Counter n TC.
The two software-selectable output options are pulse output on TC and
toggle output on TC. The output polarity is software-selectable for both
options.
With pulse or pulse train generation tasks, the counter drives the pulse(s) on
the Counter n Internal Output signal. The Counter n Internal Output signal
can be internally routed to be a counter/timer input or an “external” source
for AI, AO, DI, or DO timing signals.
Routing Counter n Internal Output to an Output
Terminal
You can route Counter n Internal Output to any output PFI terminal.
Frequency Output Signal
The Frequency Output (FREQ OUT) signal is the output of the frequency
output generator.
Routing Frequency Output to a Terminal
You can route Frequency Output to any output PFI terminal.