Routing Signals to Counter n HW Arm Input

Any of the following signals can be routed to the Counter n HW Arm input:

(USB-6210/6211/6215 Devices)

PFI <0..3>

(USB-6212/6216 Devices)

PFI <0..15>

(USB-6218 Devices)

PFI <0..3>, PFI <8..11>

AI Reference Trigger (ai/ReferenceTrigger)

AI Start Trigger (ai/StartTrigger)

Counter 1 Internal Output can be routed to Counter 0 HW Arm. Counter 0
Internal Output can be routed to Counter 1 HW Arm.

Some of these options may not be available in some driver software.

Counter n Internal Output and Counter n TC Signals

The Counter n Internal Output signal changes in response to Counter n TC.

The two software-selectable output options are pulse output on TC and
toggle output on TC. The output polarity is software-selectable for both
options.

With pulse or pulse train generation tasks, the counter drives the pulse(s) on
the Counter n Internal Output signal. The Counter n Internal Output signal
can be internally routed to be a counter/timer input or an “external” source
for AI, AO, DI, or DO timing signals.

Routing Counter n Internal Output to an Output
Terminal

You can route Counter n Internal Output to any output PFI terminal.

Frequency Output Signal

The Frequency Output (FREQ OUT) signal is the output of the frequency
output generator.

Routing Frequency Output to a Terminal

You can route Frequency Output to any output PFI terminal.