The differential connection configuration allows the common-mode noise
voltage, V

cm

, to be rejected during the measurement of V

1

.

You must connect the negative lead of your sensors and AI GND to a local
ground signal on your system.

Taking Referenced Single-Ended (RSE) Measurements

Using the RSE measurement configuration allows the
USB-6215/6216/6218 to take measurements on all AI channels when all
channels share a common ground. Figure 4-25 shows an RSE connection
configuration.

Note

If you leave the AI GND pin unconnected, the signals float outside the working input

range of the USB-6215/6216/6218. This can result in unreliable measurements because
you cannot ensure that the input signal is within 10 V of AI GND.

Figure 4-25. Connecting to the USB-6215/6216/6218 in RSE Mode

In an RSE connection configuration, each input channel is measured with
respect to AI GND.

AI 1

AI 2

AI GND

USB-6215/6216/6218

Mux

ADC

PGIA