Chapter 4

Analog Input

NI USB-621x User Manual

4-32

ni.com

difference can appear as measurement error. Follow the connection
instructions for grounded signal sources to eliminate this ground potential
difference from the measured signal.

When to Use Differential Connections with
Ground-Referenced Signal Sources

Use DIFF input connections for any channel that meets any of the
following conditions:

The input signal is low level (less than 1 V).

The leads connecting the signal to the device are greater than 3 m
(10 ft).

The input signal requires a separate ground-reference point or return
signal.

The signal leads travel through noisy environments.

Two analog input channels, AI+ and AI–, are available.

DIFF signal connections reduce noise pickup and increase common-mode
noise rejection. DIFF signal connections also allow input signals to float
within the common-mode limits of the NI-PGIA.

Refer to the

Using Differential Connections for Ground-Referenced Signal

Sources

section for more information about differential connections.

When to Use Non-Referenced Single-Ended (NRSE)
Connections with Ground-Referenced Signal
Sources

Only use non-referenced single-ended input connections if the input signal
meets the following conditions:

The input signal is high-level (greater than 1 V).

The leads connecting the signal to the device are less than 3 m (10 ft).

The input signal can share a common reference point with other
signals.

DIFF input connections are recommended for greater signal integrity for
any input signal that does not meet the preceding conditions.