beautypg.com

Analog input ground-reference settings, Analog input ground-reference settings -3, Analog input – National Instruments Data Acquisition Device NI USB-621x User Manual

Page 36: Ground-reference settings

background image

Chapter 4

Analog Input

© National Instruments Corporation

4-3

NI USB-621x User Manual

For more information about setting ranges, refer to the NI-DAQmx Help or
the LabVIEW Help in version 8.0 or later.

The following table shows the input ranges and resolutions supported by
USB-621x devices.

Analog Input Ground-Reference Settings

USB-621x devices support the following analog input ground-reference
settings:

Differential mode—In DIFF mode, USB-621x devices measure the
difference in voltage between two AI signals.

Referenced single-ended mode—In RSE mode, USB-621x devices
measure the voltage of an AI signal relative to AI GND.

Non-referenced single-ended mode—In NRSE mode, USB-621x
devices measure the voltage of an AI signal relative to the AI SENSE
input.

The AI ground-reference setting determines how you should connect your
AI signals to the USB-621x device. For more information, refer to one of
the following sections depending on your device:

Connecting Analog Input Signals on USB-6210/6211/6212 Devices

Connecting Analog Input Signals on USB-6215/6216/6218 Devices

Ground-reference settings are programmed on a per-channel basis.
For example, you might configure the device to scan 12 channels—four
differentially-configured channels and eight single-ended channels.

USB-621x devices implement the different analog input ground-reference
settings by routing different signals to the NI-PGIA. The NI-PGIA is a
differential amplifier. That is, the NI-PGIA amplifies (or attenuates) the
difference in voltage between its two inputs. The NI-PGIA drives the ADC

Input Range

Nominal Resolution Assuming

5% Over Range

–10 V to 10 V

320

μV

–5 V to 5 V

160

μV

–1 V to 1 V

32

μV

–200 mV to 200 mV

6.4

μV