beautypg.com

Figure 11-1. falling-edge trigger, Triggering with an analog source, Figure 11-2. analog trigger circuitry – National Instruments Network Device DAQ S User Manual

Page 134: Triggering with an analog source -2, Sectio

background image

Chapter 11

Triggering

NI 6124/6154 User Manual

11-2

ni.com

Figure 11-1 shows a falling-edge trigger.

Figure 11-1. Falling-Edge Trigger

You can also program your DAQ device to perform an action in response to
a trigger from a digital source. The action can affect the following:

analog input acquisitions

analog output generation

counter behavior

Triggering with an Analog Source

(NI 6124 Only)

Some S Series devices can generate a trigger on an analog

signal. Figure 11-2 shows the analog trigger circuitry.

Figure 11-2. Analog Trigger Circuitry

You must specify a source and an analog trigger type. The source can be
any analog input channel.

5 V

0 V

Falling edge initiates acquisition

Digital Trigger

Analog
Trigger

Detection

Mux

ADC

ADC

Analog Comparison Event

(Analog Trigger Circuitry Output)

AO Circuitry

AI Circuitry

Counter Circuitry

ADC

ADC

PGIA

AI 0

+

PGIA

AI 1

+

PGIA

AI 2

+

PGIA

AI 3

+