Lifescan One Touch Basic Plus Diabetes Monitoring System User Manual

Page 62

background image

MESSAGE

PROBLEM

WHAT TO DO

error3
retest

not
enough
blood
retest

The test strip was
removed during a test or
the test strip was inserted
with blood or control
solution already applied.

1. The blood or control

solution sample was
too small or smeared.

2. The test strip was not

inserted far enough
into the test strip
holder.

Repeat the test with a
new test strip. Insert the
test strip into the meter
and wait for the words
apply sample to appear.
Apply blood or control
solution with test strip
inserted in meter.

1. Repeat the test with

a new test strip and
a large, shiny drop
of blood or control
solution.

2. Repeat the test with a

new test strip, pushing
the test strip all the way
into the test strip
holder.

55