beautypg.com

Lifescan One Touch Basic Plus Diabetes Monitoring System User Manual

Page 60

background image

53

MESSAGE

PROBLEM

WHAT TO DO

If the word contrl
appears after a blood
test, your blood sample
was too small, smeared, or
another drop was added
after the test began.

The sample was applied
before the words apply
sample appeared.

1. The test strip moved

during the test.

2. The test strip was not

inserted correctly.

3. The test strip was

removed before the
test was completed.

Repeat the test with a new
test strip and a larger drop
of blood.

Repeat the test with a new
test strip.

1. Repeat the test with a

new test strip.

2. Repeat the test with a

new test strip.

3. Repeat the test with a

new test strip.

(Example)

e r r o r 1
retest

e r r o r 2
retest