Cleaning the Crossover Assemblies
Clean the crossover assemblies in the processor with a soft cloth to remove any
chemical residue on them. Install the splash guard see “Preventing Algae (for
Wash Configuration)” on page 4-8. Rinse the gears and socks with warm water
from a squirt bottle to dissolve deposits.
Operate the drive for the processor for at least 30 minutes before making prints or
running a control strip. This will ensure that the socks are thoroughly soaked with
chemistry.