Kodak RP 50 2J0686 User Manual
Page 115
Diagnostics and Troubleshooting
October 2002
5-25
1101
Density DMIN/DMAX points
are invalid.
Density points are invalid.
Reread density values on
densitometer or make a new
calibration print.
1102
Density data is not
monotonic increasing.
Density data is invalid.
Reread density values on
densitometer or make a new
calibration print.
1103
Exceeded maximum number
of allowable data errors.
Density data is invalid.
Reread density values on
densitometer or make a new
calibration print.
1104
Asymmetric Density
Groupings are invalid.
Density data is invalid.
Reread density values on
densitometer or make a new
calibration print.
1201
Unknown response from
densitometer.
The application received an
unknown response from the
densitometer during
initialization.
Verify that the baud rate of the
densitometer matches the
applications configuration.
1202
Error:_error_from
densitometer.
Unexpected error was returned
from the densitometer.
Calibrate the densitometer. If the
problem continues, check the
manual for the densitometer.
1203
Error: Number of patches
seen by densitometer is
incorrect.
Unable densities on densitometer.
Try adjusting frame counts or
reread the density patches on the
densitometer.
1300
Values have not been loaded
Matrix file missing.
Create a new device.
1301
Data or Matrix files have not
yet been loaded
Matrix file missing.
Create a new device.
1302
Data cannot be saved, must
first apply the matrix
Matrix file missing.
Create a new device.
Error
Code
Error Code /Status
Message
Possible Cause/Subsystem
Possible Solution