Delay trigger (option), Delay trigger (option) -25 – RIGOL MSO/DS1000Z Series User Manual
Page 99
Chapter 5 To Trigger the Oscilloscope
RIGOL
MSO1000Z/DS1000Z User’s Guide
5-25
Delay Trigger (Option)
In delay trigger, you need to set signal source A and signal source B. The oscilloscope
triggers when the time difference (△T) between the specified edges of source A
(Edge A) and source B (Edge B) meets the preset time limit, as shown in the figure
below.
Note:
Edge A and Edge B must be neighbouring edges.
T
Source A
Source B
Edge A=
Edge B=
Figure 5-10 Delay Trigger
Trigger Type:
Press Type to open the trigger type list. Rotate
to select "Delay" and press down
. At this point, the trigger setting information is displayed at the upper right corner
of the screen.
For example,
. The trigger type is delay trigger; the current
trigger source is CH1; the trigger level is 0.00 V.
Source A:
Press SourceA to select CH1-CH4 as the trigger source of signal source A. For the
details, please refer to the introduction in "
Trigger Source". The current trigger
source is displayed at the upper right corner of the screen.
Note:
Select channel with signal input as trigger source to obtain stable trigger.
Edge A:
Press EdgeA to select the trigger edge type of signal source A in delay trigger. It can
be set to the rising edge or falling edge.
Source B:
Press SourceB to select CH1-CH4 as the trigger source of signal source B. For the
details, please refer to the introduction in "Trigger Source". The current trigger
source is displayed at the upper right corner of the screen.
Note:
Select channel with signal input as trigger source to obtain stable trigger.