All measurement, Statistic function, All measurement -31 – RIGOL MSO/DS1000Z Series User Manual
Page 145: Statistic function -31

Chapter 6 MATH and Measurement
RIGOL
MSO1000Z/DS1000Z User’s Guide
6-31
All Measurement
All measurement could measure all the time and voltage parameters (each
measurement source has 29 items) of the current measurement source and display
the results on the screen. Press Measure Measure All to enable or disable the
all measurement function. Press All Measure Source and use
to select the
channel(s) to be measured (CH1-CH4 and MATH).
When all measurement is enabled, "One-key" measurement is also valid.
"To Clear the Measurement" will not clear the results of all measurement.
Statistic Function
Make statistic and display the current, average, minimum (or standard deviation) and
maximum (or count) values of at most 5 measurement items that are turned on last.
Press Measure Statistic to turn the statistic function on or off. When the
statistic function is enabled, press Stat.Sel. to select "Extremum" or "Difference"
measurement. When "Extremum" is selected, minimum and maximum values are
displayed. When "Difference" is selected, standard deviation and count values are
displayed. Press Measure Reset Stat. to clear the history data and make
statistic again.
Extremum Measurement
Difference Measurement