1 using the gas unit configurator tool, Igurator utility (section 5.3.1), Caution – Emerson Network Router User Manual
Page 52
46
ProLink
®
II Software for Micro Motion
®
Transmitters
Transmitter Configuration, Characterization, and Calibration
5.3.1
Using the Gas Unit Configurator tool
Note: For Model 2400S transmitters, and for Series 1000/2000 and LF-Series transmitters with
firmware version 5.0 and above, the Gas Unit Configurator tool is not available because standard
and normal units are pre-defined in the transmitter.
For many gas applications, standard or normal volume flow rate is used as the quasi mass flow rate.
Standard or normal volume flow rate is calculated as the mass flow rate divided by the density of the
gas at a reference condition.
To configure a mass flow special unit that represents standard or normal volume flow rate, you must
calculate the mass flow conversion factor from the density of the gas at a reference temperature,
pressure, and composition. ProLink II v1.2 and later offers a Gas Unit Configurator tool to calculate
this mass flow conversion factor. The tool will automatically update the mass flow conversion factor
in the
Special Units
tab.
Note: Micro Motion recommends that you do not use the meter to measure actual volume flow of a
gas (volumetric flow at line conditions). If you need to measure actual volume flow of a gas, contact
Micro Motion customer support.
To use the Gas Unit Configurator:
1. Choose
Tools > Gas Unit Configurator
.
2. Select the
Time Unit
that your special unit will be based on.
3. Select either
English Units
or
SI Units
.
4. Click
Next
.
5. Define the standard density to be used in calculations.
•
To use a fixed standard density, click the top radio button, enter a value for standard
density in the
Standard Density
box, and click
Next
.
•
To use a calculated standard density, click the second radio button and click
Next
. Then
enter values for
Reference Temperature
,
Reference Pressure
, and
Specific Gravity
on
the next panel, and click
Next
.
6. Check the values displayed.
•
If they are appropriate for your application, click
Finish
. The special unit data will be
written to the transmitter.
•
If they are not appropriate for your application, click
Back
as many times as necessary to
return to the relevant panel, correct the problem, then repeat the above steps.
CAUTION
The meter should not be used for measuring the actual volume of gases.
Standard or normal volume is the traditional unit for gas flow. Coriolis meters
measure mass. Mass divided by standard or normal density yields standard or
normal volume units.