Figure 3-2). until you connect to a transm – Emerson Network Router User Manual
Page 33
Installation and Use Manual
27
Using ProLink II Software
Pr
oL
in
k II
Setup
T
ransmitter Star
tup
Usi
ng
Pr
oL
in
k
II
Be
fo
re
Y
o
u
Begin
Figure 3-1
Connect to Densitometer/Viscometer window
Making the connection – all other transmitters
1. Ensure that your PC is connected to a transmitter, using one of the methods described in
2. Choose
Connection > Connect to Device
. See Figure 3-2.
3. Select the protocol to use. Depending on the
Protocol
option that you choose, different
communications options will be available for configuration.
4. Specify
Baud Rate
,
Parity
,
Stop Bits
,
COM Port
, and
Address/Tag
as appropriate for your
connection and transmitter. See Section 2.6 for more information on these parameters.
5. If you are using the USB HART Interface, select
Converter Toggles RTS
.
6. Click
Connect
.
Alternately, if you do not know the address of your transmitter:
1. Click
Poll
. ProLink II will poll the network for all Micro Motion transmitters, and display a
list of all transmitters found.
2. Select the transmitter to connect to, and click
OK.
Figure 3-2
Connection window (all other transmitters)