once the number stops moving. Press the on/off 
button to accept the level. The screen will briefly flash 
to acknowledge the entry. 
Switch off the synthetic/clean air and disconnect the 
clean air supply. 
Using your Cub
Press the on/off button to initiate the Span stage The 
screen will change and display an ‘S’. At this points 
connect the Span gas supply. Switch on the gas 
supply and then press the on/off button. 
The screen will change and display a live reading 
from the sensor. IMPORTANT: This number does not 
relate to gas concentration, it simply indicates if the 
sensor output is moving. The output should increase, 
once the number stops moving press the on/off 
button to accept the level. The screen will briefly flash 
to acknowledge the entry. Switch off the 
synthetic/clean air and disconnect. 
If there is not a significant difference between the 
Zero and Span levels the calibration will be invalid 
and E- - 5 will be displayed. If this occurs the previous 
valid calibration levels will be used. 
At the end of the calibration sequence the Cub will 
revert to the general VOC / normal running. 
IMPORTANT: Always check the calibration in normal 
running before use. Apply the same Zero and SPAN 
gas used to calibrate and ensure the correct readings 
are displayed. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 