Ii - 2 installation set u p – HEIDENHAIN ND 522 User Manual

Page 76

background image

76

II Technical Information

II

-

2

Installation

Set

u

p

Configuring the Compensation Table

Press the EDIT TABLE soft key to view the table entries.

Use the UP, or DOWN ARROW keys, or the numeric keys to move

the cursor to the correction point to be added, or changed. Press
Enter

.

Enter the known error which exists at this point. Press Enter.

When completed, press C key to exit the table, and return to the

ERROR COMPENSATION form.

Reading the Graph
The error compensation table may be viewed in table, or graphical

formats. The graph shows a plot of a translation error vs. measured

value. The graph has a fixed scale. As the cursor is moved through the

form, the location of the point on the graph is indicated with a vertical

line.

Viewing the Compensation Table

Press the EDIT TABLE soft key.

To switch between the table, and graph views, press the VIEW soft

key.

Press the UP, or DOWN ARROW keys, or the numeric keys to move

the cursor within the table.

The error compensation table data may be saved to, or loaded from a

PC via the USB port.

Exporting the Current Compensation Table

Press the EDIT TABLE soft key

Press the IMPORT/EXPORT soft key.

Press the EXPORT TABLE soft key.

Importing a New Compensation Table

Press the EDIT TABLE soft key.

Press the IMPORT/EXPORT soft key.

Press the IMPORT TABLE soft key.

Pressing Enter saves the Error Compensation data

entered, and is retained if the power is cycled. Pressing the

C key to exit without pressing the Enter key only retains

data until power is cycled.