beautypg.com

Enable/disable ref function – HEIDENHAIN ND 522 User Manual

Page 22

background image

22

I Operating Instructions

I -

2

Gener

al

Oper

ations

fo

r

ND

522/523

ENABLE/DISABLE REF function
The toggling ENABLE/DISABLE soft key, that is present during the

Reference Mark Evaluation routine, allows the operator to select a

specific Reference Mark on an encoder. This is important when using

encoders with Fixed Reference Marks. When the DISABLE REF soft

key is pressed, the evaluation routine is paused, and any reference

marks that are crossed during encoder movement are ignored. When

the ENABLE REF soft key is then pressed, the evaluation routine once

again becomes active, and the next crossed reference mark will be

selected.
Once reference marks for all desired axes are established, press

NO REF soft key to cancel out of routine. Only the axes that are

needed require crossing over the reference marks. If all reference

marks have been found the ND 522/523 will return to the DRO display

screen automatically.

For everyday power up; turn on power, and press any
key.

Cross over the reference marks (in any order).

Press DISABLE REF soft key, and cross over
reference marks.

Move encoder to desired fixed reference mark. Press
ENABLE REF soft key, and cross over reference
mark.

Do not cross over the reference marks, and press the
NO REF soft key. Note: In this case the relationship
between axis slide position, and display value will be
lost after a power interruption.

If the reference marks are not crossed over, the

ND 522/523 does not store the datum points. This means

that it is not possible to re-establish the relationship

between axis slide positions, and display values after a

power interruption (switch-off).

- ALTERNATIVE METHOD -

- ALTERNATIVE METHOD -

This manual is related to the following products: