Testing procedure, Unpacking and charging the cat-100, Entering the coil factor – Super Systems CAT-100 User Manual
Page 8

CAT-100 Operations Manual
Super Systems Inc.
Page 8 of 51
Testing Procedure
Unpacking and Charging the CAT-100
Carefully remove the packing tape from the box in which the CAT-100 was shipped and remove
the CAT-100, right side up. Using proper lifting technique, place the CAT-100 on a level surface.
If you are going to use the CAT-100 without electrical outlet power, allow the unit to charge
overnight before use. Do this by simply plugging the unit into an AC electrical outlet and
allowing it to charge. Make sure that the unit is charging with the power turned OFF.
Entering the Coil Factor
The Coil Factor for the wire coil is used in calculating Carbon Potential and is based on
variations in coils in a batch. The recommended Coil Factor is printed on each bag of coils and is
the same for all coils in a batch. Before using the coil in the testing process, follow this
procedure for entering and verifying the Coil Factor. It is essential that this procedure be
followed correctly.
First, note the recommended Coil Factor printed on the bag of coils that the coil you are using
came from. If you do not have the bag, assume a Coil Factor of 100.
Next, log in with Configuration Mode
on the CAT-100. This is done by
opening the “LOGIN” screen from the
main menu and entering the
Configuration pass code (which is ‘2’
by default).
Figure 3 – Entering Configuration pass code (default is ‘2’)
Open the Detail menu from the main
menu. The Detail screen will be
shown, and the Coil Factor will be
displayed in a box labeled “Factor”.
Tap on that box.
Figure 4 – Detail screen with Coil Factor box