beautypg.com

Troubleshooting – Super Systems CAT-100 User Manual

Page 47

background image

CAT-100 Operations Manual

Super Systems Inc.

Page 47 of 51

Troubleshooting

The CAT-100 is designed to provide accurate results testing procedures and conditions are consistent with the product’s design. When

problems or discrepancies are encountered, they can often be corrected. The table below provides a list of potential problems that may be

encountered, possible causes, and corrective actions that may rectify the problem.


WARNING!

When troubleshooting, follow all proper safety precautions. Use proper eye protection and hand protection at all times.


Problem

Possible Causes

Possible Corrective Actions

Wire breaks at any point in the procedure

Coil cooled too quickly

Wire stretched beyond design limitations

When removing insertion rod, follow procedure

exactly as shown in Cooling Down and Removing

the Insertion Rod

section on page 14

Avoid stretching wire excessively

% Carbon indicated by CAT-100 differs by more

than 0.03% from Process Variable shown by

instrumentation at the time wire was exposed to

testing atmosphere

Wire soaked too early in soak cycle

Emery cloth not applied to wire after soaking

Variability in % Carbon levels in testing

atmosphere

Testing procedure not followed correctly

Instrumentation error

Insert wire coil into testing atmosphere during

final one-third of soak cycle.

Perform a test of a new wire, closely following

the instructions in Using the Emery Cloth on the

Steel Wire Coil

page on 15

Ensure that testing atmosphere remains

consistent during test. Temperatures, % Carbon,

and other parameters should not change

significantly while wire is exposed to testing

environment. Furnace should be in a soak state

for a minimum of one hour.

Calibrate CAT-100 and re-test using an unused

wire, following instructions exactly as indicated in

manual

Contact equipment manufacturer or distributor