Login, Calibration, Login calibration – Super Systems CAT-100 User Manual
Page 23: Ee login, E 23), Figure 32)
CAT-100 Operations Manual
Super Systems Inc.
Page 23 of 51
Figure 32 - About Screen
LOGIN
The purpose of the LOGIN option (Figure 33) is to allow the user to access Supervisor and
Configuration functions. Enter the Supervisor pass code (default is 1) to log in with the
Supervisor access level. Enter the Configuration pass code (default is 2) to log in with the
Configuration pass code. Functions available with each access level are shown in the CAT-100
Menu Navigation section on page 19.
Figure 33 - LOGIN Option
Calibration
A circuit calibration is required once per year and recommended every four months. The
required annual calibration is performed by SSi, and a certificate is issued for this calibration. A
circuit calibration can be performed by the end-user by using the Calibration Bridge provided by
SSi when the CAT-100 is first shipped.
An important guideline is as follows: A circuit calibration should be performed at any time the
“Error in %” found in this procedure is greater than 0.5 or less than -0.5.
NOTE: Information on ordering a replacement Calibration Bridge can be found in the
Replacement Parts section (page 6).