Problem
Possible Causes
Possible Corrective Actions
% Carbon indicated by CAT-100 differs by more
than 0.03% from Process Variable shown by
instrumentation at the time wire was exposed to
testing atmosphere
(Continued)
Incorrect Coil Factor entered
Ensure that the correct Coil Factor is entered
during the testing procedure; see Entering the
Coil Factor section on page 8.
Carbon Potential is not within the effective range
of 0.1% to 1.3%
Actual Carbon Potential is not within the effective
range of the unit
Soak time too long or too short
% Carbon shown must be within the effective
range in order to be reliable. A different method
of measuring Carbon Potential may be needed in
such a case.
Re-test using an unused wire. Ensure that soak
times shown in Table 2 are applied as indicated.
Wire becomes detached from insertion rod
during soak in testing atmosphere
Unsecured attachment
Before inserting rod into testing atmosphere,
verify that wire coil is in place securely for
testing, that rod screw is securely tightened, and
that protective cage is securely in place on the
rod
Reading is not stored when “Store” is pressed on
CAT-100 touch screen
Pressure on screen not hard enough
Interference from protective film over screen
No reading to store
Internal storage error
Increase pressure of contact with screen. Take
care not to cause damage to screen or protective
film.
Verify that spacing between protective film and
touch screen allows for contact with touch screen
Verify that a test result was generated when wire
coil was connected to binding posts on CAT-100.
If not, reconnect wire to binding posts and
observe reading again.
Call SSi Technical Support at
(513) 772-0060
Test results not displaying correctly in Test
Results screen
Internal storage error
Call SSi Technical Support at
(513) 772-0060