Specification

43

Flash Test (CLASS 1)

Test

Voltage

1.25kV

a.c

Display Range

0.1 - 10 mA a.c.

Resolution

0.01mA

Accuracy

+/- 5% +/- 2 digits

Pass Levels

selectable up to 10mA

Flash Test (CLASS2)

Test

Voltage

2.5kV

a.c

Display Range

0.1 - 10 mA a.c.

Resolution

0.01mA

Accuracy

+/- 5% +/- 2 digits

Pass Levels

selectable up to 10mA

Leakage

Test

Voltage

Mains

Supply

Display Range

0.1 - 20.0 mA a.c.

Resolution

0.01mA

Accuracy

+/- 10% +/- 2 digits (1.00 - 20mA)

Pass Levels

selectable up to 10mA

Power Tests

Resolution

0.01kVA

Measured Load

0 - 4kVA

Accuracy

+/- 10 of reading, +/- 2 digits

IEC Lead Test

Test

Voltage

40Vac

Detects

Good,

Open,

Short,

Reversed

Mechanical

Size

450mm x 410mm x 155mm

Weight

3.5kg