Seaward 220 PrimeTest User Manual
Page 29
Operating Instructions
Primetest 220
22
9.4.8 Insert a new set of batteries into the battery
compartment ensuring that the battery polarity
matches the marking on the inside of the battery
compartment.
9.4.9 Relocate the battery cover over the battery
compartment and fasten in position with the
battery cover captive screw.
This meter contains NiMH batteries. Do not
dispose of these batteries with other solid
waste. Used batteries should be disposed of by
a qualified recycler or hazardous materials
handler.
9.6 Checking the EUT Socket Fuse.
9.6.1 Plug the PrimeTest 220 into the mains power
socket, using the mains lead provided.
9.6.2 Select one of the four power tests, protective earth
conductor current ,touch leakage current ,
RCD trip or earth loop impedance test
Zs. on
the rotary switch 11
9.6.3 If the fuse is intact then the symbols PE, PN or NE
will be displayed in the appropriate manner (see
section 8.2.4) depending on the configuration of
the power socket being used.
9.6.4 If the fuse is defective then the symbols PE, PN and
NE will all flash on the display.
9.7 Replacing the Fuse.
Before opening the PrimeTest 220 ensure that it
is disconnected from all voltages!
Electric shock danger!
The replacement fuse type is specified for
ratings and size on the battery compartment
cover on the rear of the PrimeTest 220.