Operating Instructions

Primetest 220

16

8.4.9 An insulation resistance that fails will terminate the

lead test, a pass will automatically sequence to the
lead wiring test and the small digital display 3 will
indicate “LEAD”.

8.4.10 The wiring of the IEC lead is tested over a period of

2 seconds. The large digital display 2 indicates a
PASS/FAIL status for the lead. An additional visual
indication of a ✓ or ✗ also indicates PASS/FAIL
status.

The IEC lead wiring test checks for open circuit,

short circuit or crossed wires within the IEC

lead.

8.4.11 If this is the last IEC lead test or the only test then

the results will remain displayed until another rotary
switch or push button action.

8.5 Safety Test, Class II

See Figure 3

8.5.1 Select the Class II test function on the rotary

switch 10 . The LCD will display a probe symbol
to indicate that a safety test lead is required.
Ensure the IEC mains lead is disconnected from the
PrimeTest 220.

8.5.2 Plug the equipment under test (EUT) into the EUT

socket 11 on the front of the PrimeTest 220.

8.5.3 Plug the black insulation test lead into the black,

socket 12 on the PrimeTest 220.

8.5.4 Connect the other end of the black insulation test

lead to any exposed metal parts on the outer cover
of the EUT or near the mains input or mains switch
if there are no exposed metal parts.

8.5.5 To start a Class II safety test, press the button

9 .