Operating Instructions

Primetest 220

5.6 Dimensions of the PrimeTest 220 are 265 x 108 x

55mm (L x W x H)

6.0 Environmental Conditions

6.1 The PrimeTest 220 has been designed to perform

tests and measurements in a dry environment.

6.2 Maximum barometric elevation for making

measurements is 2000M.

6.3 Overvoltage category IEC 60664/IEC 61010, 300V

Category II.

6.4 Contamination degree 2 according to IEC 61010-1.

6.5 Protective system IP40 according to IEC 60529.

6.6 Electromagnetic compatibility (EMC). Interference

immunity and emitted interference conforming to
IEC 61326-1.

6.7 Operating temperature range of 0°C to 40°C,

without moisture condensation.

6.8 The PrimeTest 220 can be stored at any

temperature in the range -25°C to +65°C (relative
humidity up to 90%). The batteries should be taken
out of the instrument for storage.

7.0 Electrical Specification

7.1 Earth continuity measuring range (rotary switch

position , and IEC)

Measuring Range Resolution

Accuracy Overload Protection

19.99

0.01

±5% ±2 digits

300 V AC/DC

Open circuit voltage

>4VDC

Test current

>220mA (0-2

)

The maximum number of continuity tests, with a duty
cycle specified by BS EN 61557-4 Section 6.6, is 1900
repetitions.