Operating Instructions
Primetest 220
8
7.2 Insulation resistance measuring range (rotary
switch position , , IEC and M
Ω
.
Measuring Range Resolution
Accuracy Overload Protection
2M
Ω
0.01M
Ω
±2% ±2 digits
300 V AC/DC
20M
Ω
0.1M
Ω
±5% ±2 digits
300 V AC/DC
200M
Ω
1M
Ω
±10% ±2 digits
300 V AC/DC
Test voltage
500VDC at 1mA,
<750VDC for o/c
Test voltage accuracy
+20%, -0%
Test current
<2mA for s/c
The maximum number of insulation tests, with a duty
cycle specified by BS EN 61557-2 Section 6.7, is 2000
repetitions.
7.3 IEC lead test
The IEC lead test is a sequence of earth continuity
measurement (section 7.1) - insulation resistance
measurement (section 7.2) - Live/Neutral checks for o/c,
s/c and crossed conductors in the IEC lead.
Test voltage
9 VDC
Test duration
10 seconds
7.4 Touch current measurement
Measuring Range Resolution
Accuracy
Overload Protection
3.5mA
0.01mA ±10% ±2 digits (5° - 40°)
N/A
±15% ±2 digits (0° - 5°)
Test voltage
mains supply @ 230V -
15%/+10% @ 10A max
Frequency
50Hz ±1%
Test duration
2 seconds minimum
Frequency response
IEC 61010-1 Annex A.1
The test is disabled when phase-neutral (PN) or phase-
earth (PE) voltage is not at mains potential or a voltage of
40 V AC or greater exists between neutral-earth (NE).