beautypg.com

Seaward 220 PrimeTest User Manual

Page 18

background image

Operating Instructions

Primetest 220

11

During an insulation resistance measurement a

high voltage is applied between the Live/Neutral

and Earth of the EUT. It should be noted that

this voltage can also be present on the bare

metalwork of the EUT.

8.1.10The insulation resistance of the EUT is measured

over a period of 2 seconds. The large display 2
indicates digital insulation resistance, M

, the

small display 3 indicates a PASS/FAIL status. An
additional visual display of a ✓ or a ✗ also indicates
PASS/FAIL status.

8.1.11The insulation resistance value will remain

displayed until another rotary switch action.

8.2 Protective Earth Conductor Current

See figure 4.

Attention - The Black 4mm test socket must not

be used whilst the PrimeTest 220 is performing

a leakage test. Electric Shock Danger.

Attention - before a protective earth conductor

current test is performed on a Class I EUT the

operator must ensure that the EUT has already

passed a Class I safety test.

Attention - An RCD may be triggered if the EUT

is defective.

Warning - it is important that the user verifies

that an appliance with moving parts (e.g. an

electric drill) is safely mounted to allow

movement without causing damage to

equipment or personnel.

8.2.1 Plug the supplied mains power lead into the mains

power socket 13 on the PrimeTest 220 .

8.2.2 Plug the other end of the supplied mains power

lead into a mains power socket.

PrimeTest 220 REV 1 5/12/05 1:52 pm Page 16