Seaward 220 PrimeTest User Manual
Page 16
Operating Instructions
Primetest 220
9
7.5 Protective earth conductor current
measurement
Measuring Range Resolution
Accuracy Overload Protection
19.99mA
0.01mA ±5% ±2 digits
N/A
Test voltage
mains supply @ 230V -
15%/+10% @ 10A max
Frequency
50Hz ±1%
Test duration
2 seconds minimum
Frequency response
IEC 61010-1 Annex A.1
(from 40Hz)
The test is disabled when phase-neutral (PN) or phase-
earth (PE) voltage is not at mains potential or a voltage of
40 V AC or greater exists between neutral-earth (NE).
7.6 RCD test
Test current
30mA r.m.s
Test current accuracy
+5%, - 0%
Test duration
500ms maximum if RCD
does not trip
Trip time accuracy
0mS to 40mS, ±2mS
41mS to 500mS, ±5mS
The test is disabled when phase - neutral (PN) or phase
-earth (PE) voltage is not at mains potential or a voltage
of 40 VAC or greater exists between neutral - earth (NE).
7.7 Zs test
Measuring Range
Resolution
Accuracy Overload Protection
19.9
Ω
0.1
Ω
±10% ±2 digits
N/A
199
Ω
1
Ω
±10% ±2 digits
N/A
Test current
Such that it will not trip a
30mA RCD
Test voltage
230VAC +10%, -15%
The test is disabled when phase-neutral (PN) or phase-
earth (PE) voltage is not at mains potential or a voltage of
40VAC or greater exists between neutral-earth (NE).
PrimeTest 220 REV 1 5/12/05 1:52 pm Page 14