Operating Instructions

Primetest 220

19

8.6.11 The touch leakage current will remain displayed

until another rotary switch or pushbutton action.

8.7 Earth Loop Impedance, Zs, Test

See Figure 9.

8.7.1 Plug the supplied mains power lead into the mains

power socket 13 on the PrimeTest 220.

8.7.2 Plug the other end of the supplied mains power

lead into a mains power socket.

8.7.3 Select the

Zs test function on the rotary switch 10

8.7.4 Switch on the mains power.

8.7.5 The PrimeTest 220 automatically checks the wiring

of the power socket being used. If the line
potentials are correct then the LCD will display the
symbols PE and PN in the steady state. Where a
fault exists then the appropriate symbol will flash
(see Appendix 1) along with the symbol being
displayed in the steady state. A fault condition will
inhibit the

Zs test.

If a power test is selected but the mains supply

has not been connected to the PrimeTest 220

then PE and PN symbols will flash and the test

is inhibited.

8.7.6 To start an earth loop impedance test press the

button 9 .

8.7.7 Thelarge display 2 indicates a digital resistance

value, in

.

8.7.8 The earth loop impedance value will remain

displayed until another rotary switch or push button
action.

PrimeTest 220 REV 1 5/12/05 1:52 pm Page 24