beautypg.com

RJS Inspector D4000 Laser (FIRMWARE version A.05 and Earlier) User Manual

Page 76

background image

Appendix K

Operator

s Guide

RJS, Minneapolis, MN

002-5568

70

Reference Decode

Each symbology type specifies a specific decoding method to be used in determining
overall symbol grade.

This parameter grade can be “A” or “F.” (“A” is pass and “F” is fail.) If this parameter
is “F” the overall symbol grade will also be “F” regardless of any other parameter. If this
parameter is “A” the lowest of the other parameter grades determines the overall symbol
grade.

Reflectance

A measure of the amount of light reflected from an illuminated surface.


Refl(min) / Refl (max)

This parameter grade can be “A” or “F.” The reflectance value of the “lightest” space in
a symbol must be equal to or greater than twice the reflectance value of the “darkest”
bar.

Refl(max) = or > Refl(min) x 2 if “yes” “A” else “F”


Reflectance Calibration Standard

A standard or “known” reflectance value, usually printed on the back of a template or
plaque. Barcode readers and verifiers are calibrated for reflectance using these known
values.


Required PCS

This is the minimum PCS percent required for the symbol.


Resolution

The dimension of the smallest code element that can be printed; the higher the resolution
the clearer the image.

This manual is related to the following products: