change your system grounding configuration or add isolation signal conditioning. (Please look at our ISO-RACK and
ISO-5B-series products to add electrical isolation, or give our technical support group a call to discuss other options).

NOTE

Relying on the earth prong of a 120VAC for signal ground connections is not advised.. Different ground
plugs may have large and potentially even dangerous voltage differentials. Remember that the ground
pins on 120VAC outlets on different sides of the room may only be connected in the basement. This
leaves the possibility that the “ground” pins may have a significant voltage differential (especially if the
two 120 VAC outlets happen to be on different phases!)

CIO-DAS16/F and Signal Source Having Isolated Grounds

Some signal sources will already be electrically isolated from the CIO-DAS16/F. The diagram below shows a typical
isolated ground system. These signal sources are often battery powered, or are fairly expensive pieces of equipment
(since isolation is not an inexpensive proposition), isolated ground systems provide excellent performance, but require
some extra effort during connections to assure optimum performance is obtained. Please refer to the following sections
for further details

3.3 WIRING CONFIGURATIONS

Combining all the grounding and input type possibilities provides us with the following potential connection
configurations. The combinations along with our recommendations on usage are shown in Table 3-1 below.

Table 3-1. Input Type Recommendations

Recommended

Differential Inputs

Already Isolated

Grounds

Acceptable

Single-ended Inputs

Already Isolated Grounds

Unacceptable without

adding Isolation

Differential Inputs

Common Mode

Voltage > +/-10V

Unacceptable without

adding Isolation

Single-Ended Inputs

Common Mode

Voltage > +/- 10V

Recommended

Differential Inputs

Common Mode

Voltage < +/-10V

Not Recommended

Single-Ended Inputs

Common Mode

Voltage < +/-10V

Acceptable

Differential Inputs

Common Ground

Recommended

Single-Ended Inputs

Common Ground

OUR VIEW

INPUT

CONFIGURATION

GROUND

CATEGORY

The following sections depicts recommended input wiring schemes for each of the eight possible input
configuration/grounding combinations.

13