3 wiring configurations – Measurement Computing CIO-DAS16 User Manual
Page 17

change your system grounding configuration or add isolation signal conditioning. (Please look at our ISO-RACK and
ISO-5B-series products to add electrical isolation, or give our technical support group a call to discuss other options).
NOTE
Relying on the earth prong of a 120VAC for signal ground connections is not advised.. Different ground
plugs may have large and potentially even dangerous voltage differentials. Remember that the ground
pins on 120VAC outlets on different sides of the room may only be connected in the basement. This
leaves the possibility that the “ground” pins may have a significant voltage differential (especially if the
two 120 VAC outlets happen to be on different phases!)
CIO-DAS16/F and Signal Source Having Isolated Grounds
Some signal sources will already be electrically isolated from the CIO-DAS16/F. The diagram below shows a typical
isolated ground system. These signal sources are often battery powered, or are fairly expensive pieces of equipment
(since isolation is not an inexpensive proposition), isolated ground systems provide excellent performance, but require
some extra effort during connections to assure optimum performance is obtained. Please refer to the following sections
for further details
3.3 WIRING CONFIGURATIONS
Combining all the grounding and input type possibilities provides us with the following potential connection
configurations. The combinations along with our recommendations on usage are shown in Table 3-1 below.
Table 3-1. Input Type Recommendations
Recommended
Differential Inputs
Already Isolated
Grounds
Acceptable
Single-ended Inputs
Already Isolated Grounds
Unacceptable without
adding Isolation
Differential Inputs
Common Mode
Voltage > +/-10V
Unacceptable without
adding Isolation
Single-Ended Inputs
Common Mode
Voltage > +/- 10V
Recommended
Differential Inputs
Common Mode
Voltage < +/-10V
Not Recommended
Single-Ended Inputs
Common Mode
Voltage < +/-10V
Acceptable
Differential Inputs
Common Ground
Recommended
Single-Ended Inputs
Common Ground
OUR VIEW
INPUT
CONFIGURATION
GROUND
CATEGORY
The following sections depicts recommended input wiring schemes for each of the eight possible input
configuration/grounding combinations.